刘伯山,宋莉,张玉香,邹英华,赵国栋,李响,巩悦勤.运动伪影校正技术改善肾动脉瘤锥形束CT图像质量[J].中国介入影像与治疗学,2024,21(6):363-367
运动伪影校正技术改善肾动脉瘤锥形束CT图像质量
Motion freeze technology for improving quality of cone-beam CT images of renal artery aneurysms
投稿时间:2023-12-05  修订日期:2024-04-01
DOI:10.13929/j.issn.1672-8475.2024.06.011
中文关键词:  动脉瘤  伪影  锥形束计算机体层摄影术
英文关键词:aneurysm  artifacts  cone-beam computed tomography
基金项目:
作者单位E-mail
刘伯山 北京大学第一医院介入血管外科, 北京 100034  
宋莉 北京大学第一医院介入血管外科, 北京 100034  
张玉香 北京大学第一医院介入血管外科, 北京 100034  
邹英华 北京大学第一医院介入血管外科, 北京 100034 13801105222@139.com 
赵国栋 北京大学第一医院介入血管外科, 北京 100034  
李响 北京大学第一医院介入血管外科, 北京 100034  
巩悦勤 北京大学第一医院介入血管外科, 北京 100034  
摘要点击次数: 232
全文下载次数: 209
中文摘要:
      目的 观察运动伪影校正技术(MFT)对于改善肾动脉瘤锥形束CT(CBCT)图像质量的价值。方法 前瞻性选取20例拟接受经导管动脉栓塞治疗的肾动脉瘤患者,于CBCT引导下行肾动脉造影,采用MFT处理运动伪影;对MFT校正前、后图像质量进行主、客观评价,评估其显示肾动脉、载瘤动脉及靶区血管树情况,测算单位密度像素平均值(mean)、标准差(SD)及血管边缘锐化误差率。结果 MFT校正后,肾动脉瘤最大密度投影图像显示肾动脉、重建CBCT图像显示载瘤动脉均明显优于校正前(P均<0.05);利用Flightplan for Liver软件可自动提取清晰的血管树;肾动脉瘤CBCT图像SD及血管边缘锐化误差率明显减小而mean明显增加(P均<0.05)。结论 MFT可有效改善肾动脉瘤CBCT图像质量。
英文摘要:
      Objective To observe the value of motion freeze technology (MFT) for improving the quality of cone-beam CT (CBCT) images of renal artery aneurysms. Methods Twenty patients with renal artery aneurysms who would receive transcatheter arterial embolization were prospectively enrolled. Renal artery angiography was performed under the guidance of CBCT, and motion artifacts were processed using MFT. Then subjective and objective evaluation of images before and after MFT correction were performed, aimed on displaying of renal artery, parent artery and vascular tree in target area, and the mean value (mean) and standard deviation (SD) of unit density pixels, margin sharpening error rate of blood vessels were measured and calculated. Results After MFT correction, the displaying of renal artery on maximum intensity projection images and parent artery on reconstructed CBCT images were significantly better than those before correction (both P<0.05), clear vascular trees could be automatically extracted using Flightplan for Liver software, SD and margin sharpening error rate of blood vessels of renal artery aneurysm significantly decreased but mean significantly increased on CBCT images (all P<0.05). Conclusion MFT could effectively improve the quality of CBCT images of renal artery aneurysms.
查看全文  查看/发表评论  下载PDF阅读器